Measurement range |
3 % ~ 18 % |
Testing parameters |
Precision (CV) < 3 % Accuracy (V) < 1.5 % |
Sample type |
Venous blood (EDTA anticoagulation), Peripheral finger blood |
Sample volume |
5 µl whole blood for each sample |
Photometer |
415 nm LED integral flow colorimeter |
Sampling mode |
(20 samples) position 2 quality control position
1 emergency position
1 cleaning position
1 zero position |
Test mode |
Automated sample batch test Single sample test |
Calibration |
2 points or 4 points calibration, Manual or automation with a set of calibrators |
Thermostatic control |
25 ± 0.5°C (column and reagent thermostatic control) |
Reagent matching |
Column and reagent consumable / 300 pieces of testing sample |
Reagent allocation |
A, B, C, D eluent, calibrator, hemolytic agent, pump tubing |
Display |
320 x 240 LCD graphic display for real-time testing curve |
Printer |
58mm Thermal printer, printing testing curve and report |
Report output |
IFCC concentration value NGSP area percentage ADAG average glucose |
Data storage |
1000 tests report (including testing curve) |
Interface |
RS-232 interface HIS/LIS system interface |
Power supply |
AC 220V / 50Hz |
Working environment |
Temperature: 12°C ~ 30°C Relative Humidity: < 85 % |
Dimensions |
408 x 390 x 230 mm |
Weight |
9 Kgs |